Calibration intervals, pass or fail decisions with guard bands, and uncertainty budgets. Try them with your own numbers.

Budget, decision, interval

Try them with your own numbers. Everything runs in your browser, and nothing you type is sent anywhere.

SourceValue ±Distributioncu (share)Remove

Value is the stated uncertainty for Normal (k = 2), the standard deviation for Normal (k = 1), and the half-width ± a for Rectangular, Triangular and U-shaped. c is the sensitivity coefficient, usually 1.

Combined standard uncertainty
Expanded uncertainty U
Largest contributor

How long can an instrument go?

Five past calibrations of a type K thermocouple used near 900 °C show a steady drift. Set how much error you can tolerate and the chart works out when the next calibration is really due, with a safety margin built in.

18months

Suggested interval, with a 20% guard band under the limit.

Example data, invented to show the idea. A class 2 type K thermocouple at 900 °C is allowed about ±6.75 °C.

Pass, fail, or undetermined?

A temperature reading is never an exact number. When its uncertainty crosses a tolerance limit, a plain pass or fail can be wrong. A guard band keeps you from accepting, or rejecting, something you can't be sure about.

Undetermined

Pass
The value plus U is still inside the limit. Accept.
Undetermined, pass side
The value is inside the limit, but the value plus U goes over. A guard band of U does not accept it.
Undetermined, fail side
The value is over the limit, but the value minus U is back inside. It is not rejected outright.
Fail
The value minus U is already over the limit. Reject.

Example: a thermometer check where the error limit is 1.0 °C, with a guard band equal to U, the simplest rule. Real decision rules come from your standard or contract, and the wording varies, for example conditional pass and conditional fail. Example values.