Hicham Bentaous

Current calibration

Current is usually measured as the voltage across a known resistance, so its accuracy follows the shunt, the voltage measurement and heat. Clamp meters and coils bring extra traps.

Instruments and typical accuracy

InstrumentTypical rangeTypical accuracyWhere it is used
Bench multimeter, current ranges10 µA to 10 A±(0.03 to 0.1 % of reading + range term)Calibration labs, production test
Shunt resistor with voltmeter1 mA to 1000 A and more0.01 to 0.1 %, set by the shuntHigh-current reference and power measurement
Current calibratorMicroamps to 20 A±(0.01 to 0.05 %)Calibrates multimeters and panel meters
Clamp meter (Hall or transformer)1 A to several thousand A±1 to ±3 % of readingElectricians, HVAC, power systems
Transconductance amplifier with coilTo hundreds of A0.05 to 0.5 %Calibrates clamp meters

AC current adds frequency and waveform: a true RMS meter is rated for a crest factor. The accuracy figures are typical ranges to help you compare instruments. Always use the specification of the instrument in front of you.

Reference standards and traceability

  • Traceability. Current follows from voltage and resistance: I = V / R, so the standards are a voltage reference and a standard resistor or shunt.
  • Shunts. Four-terminal shunts with low power coefficient are used as transfer standards for DC and low-frequency AC.
  • Calibrators. A multifunction calibrator is the usual working standard for meter ranges below 20 A.
  • Quality system. ISO/IEC 17025, scope listing DC and AC current with frequency and range.

How it is calibrated

Typical procedure

  1. Check safety first: fuses, rated leads and lockout for the high-current ranges.
  2. Connect the meter in series with the source (or the shunt to a voltmeter for high currents).
  3. Check each range at zero, about 10 %, 50 % and 90 to 100 % of full scale in both polarities.
  4. For AC, check at the frequencies of use, for example 50 or 60 Hz and 400 Hz.
  5. For clamp meters, wind several turns through the jaw and divide by the number of turns, centring the conductor.
  6. Compare with the reference, calculate the error and report with uncertainty.

Where it goes wrong

  • Burden voltage. A meter on the 10 A range can drop a volt or more, changing the circuit current.
  • Shunt heating. Resistance drifts with power, so let the shunt settle after each change.
  • Blown fuses. Leaving a lead in the current jack and measuring voltage is the classic.
  • Clamp position. Off-centre conductors or neighbouring cables bias the reading.
  • Lead resistance. At 10 A, 10 mΩ of leads is 0.1 V.

What goes in the uncertainty budget

  • Reference standard uncertainty, including the shunt or resistor and the voltmeter.
  • Shunt temperature rise and power coefficient.
  • Burden voltage and source stability.
  • Frequency response and waveform for AC.
  • Resolution and repeatability of the unit under test.

Combine the contributions as in the uncertainty budget calculator, then decide pass or fail with a guard band as explained on the calculators page.