Hicham Bentaous

Flow calibration

Flow meters are checked against a reference that measures the same fluid: a weighed tank, a master meter or a prover. Fluid properties and the pipe around the meter matter as much as the meter.

Instruments and typical accuracy

InstrumentTypical rangeTypical accuracyWhere it is used
Coriolis mass flow meterWide, mass flow±0.05 to ±0.2 % of readingCustody transfer, batching, chemicals
Electromagnetic meterConductive liquids±0.2 to ±0.5 % of readingWater, wastewater, slurries
Ultrasonic, in-lineLiquids and gases±0.5 to ±1 % of readingWater, oil and gas pipelines
Ultrasonic, clamp-onLiquids in pipes±1 to ±3 % of readingChecks without cutting the pipe
Turbine meterClean low-viscosity fluids±0.25 to ±1 % of readingFuel, water, gas
Vortex meterSteam, gas and liquid±1 to ±1.5 % of readingProcess plants, steam
Differential pressure (orifice)Many sizes±1 to ±2 % of readingGas and steam, with ISO 5167
Positive displacement meterHigh viscosity liquids±0.1 to ±0.5 % of readingOil, fuel, dispensing
Thermal mass or rotameterGas, low flow±1 to ±5 % (rotameters of full scale)Gas dosing and indication

Accuracy usually degrades at low flow: check the turndown ratio in the specification. The accuracy figures are typical ranges to help you compare instruments. Always use the specification of the instrument in front of you.

Reference standards and traceability

  • Gravimetric method. Collect the liquid in a weighed tank for a measured time. Traceable to mass and time, with density from temperature. ISO 4185 describes it for liquids.
  • Master meters and provers. A calibrated reference meter or a piston or small-volume prover in series with the meter under test.
  • Meter standards. ISO 5167 for differential pressure devices, OIML R 117 for liquid measuring systems and ISO 17089-1 for ultrasonic meters.
  • Quality system. ISO/IEC 17025, scope listing fluid, flow range and uncertainty.

How it is calibrated

Typical procedure

  1. Stabilise liquid temperature and flow, and purge air from the line and the meter.
  2. Check the pipe: straight lengths, alignment of gaskets, full pipe and the right direction.
  3. Run at several flows, commonly 10, 25, 50, 75 and 100 % of the range, with at least three runs each.
  4. Collect for a long enough time so the start and stop timing is negligible, typically 30 to 60 s or more.
  5. Compare the meter volume or totaliser with the reference, and compute the error at each flow.
  6. Report error against flow, with the fluid, temperature and uncertainty, and adjust the K-factor or calibration factor if required.

Where it goes wrong

  • Air in the line. Bubbles read as flow in many meters and ruin the reference.
  • Poor flow profile. Elbows and valves just upstream distort the flow.
  • Wrong fluid. Viscosity and density change meter response, especially turbines.
  • Timing and diverter. A diverter that opens late adds volume to the collection.
  • Pulsation and cavitation. Pumps create pulsations that bias many meters.

What goes in the uncertainty budget

  • Reference uncertainty: weighing, density, timing or master meter.
  • Temperature and pressure of the liquid.
  • Repeatability of the meter and of the runs.
  • Installation effects and flow profile.
  • Diverter and start and stop timing.

Combine the contributions as in the uncertainty budget calculator, then decide pass or fail with a guard band as explained on the calculators page.

Calculators

Everything runs in your browser. Nothing you type is sent anywhere.

← ForcepH →