Resistivity and conductivity calibration

Resistivity covers two different jobs: solutions (conductivity cells, ultrapure water) and materials (wires, wafers, insulators). Both rest on a geometry factor and a temperature correction.

Instruments and typical accuracy

InstrumentTypical rangeTypical accuracyWhere it is used
Two-electrode conductivity cell0.01 µS/cm to about 20 mS/cm±0.5 to ±2 % of readingWater treatment, laboratory
Four-electrode or toroidal cell100 µS/cm to over 1 S/cm±0.5 to ±2 %Brines, seawater, process liquids
Ultrapure water resistivity meterUp to 18.2 MΩ·cm±1 to ±2 %Semiconductor and pharmaceutical water
Four-point probe0.001 to 1000 Ω·cm for semiconductors±1 to ±5 %Wafers, thin films, metals
Kelvin bridge or micro-ohmmeter on a sampleLow resistance of wire and bar0.1 to 1 %Wire, bus bar and conductor testing
Surface and volume resistivity test10⁶ to 10¹⁷ Ω10 to 20 % typicalInsulators, plastics, ESD materials

Resistivity is the inverse of conductivity. Pick the unit that suits the range. The accuracy figures are typical ranges to help you compare instruments. Always use the specification of the instrument in front of you.

Standard solutions and pure water at 25 °C

StandardConductivityResistivity
Ultrapure water0.0550 µS/cm18.18 MΩ·cm
0.001 mol/L KClabout 147 µS/cmabout 6.8 kΩ·cm
0.01 mol/L KClabout 1413 µS/cmabout 708 Ω·cm
0.1 mol/L KClabout 12.88 mS/cmabout 77.6 Ω·cm

Round figures for orientation. Use the certified value and temperature on your standard solution certificate.

Reference standards and traceability

  • Standard solutions. Potassium chloride solutions of known concentration reproduce conductivity at 25 °C (OIML R 56). Common values are about 147 µS/cm (0.001 mol/L), 1413 µS/cm (0.01 mol/L) and 12.88 mS/cm (0.1 mol/L): use the value on your certificate.
  • Water methods. ASTM D1125 and ISO 7888 cover electrical conductivity of water, and ultrapure water standards describe temperature-dependent limits.
  • Material methods. ASTM F84 (four-point probe for silicon), ASTM D257 (insulating materials) and IEC and ASTM wire resistivity methods.
  • Quality system. ISO/IEC 17025, with scope listing conductivity or resistivity with range and uncertainty.

How it is calibrated

Typical procedure

  1. Rinse the cell with deionised water, then with a little of the standard.
  2. Immerse in a standard that brackets the working range and wait for temperature to stabilise.
  3. Read conductance or conductivity with temperature compensation set as the standard certificate requires.
  4. Compute the cell constant, or adjust the meter, and check a second standard.
  5. For wires and bars, measure resistance with a 4-wire method, length and cross-section, and calculate ρ.
  6. For wafers, use four-point probe measurements with correction factors for size and edges.

Where it goes wrong

  • Temperature. Conductivity changes about 2 % per °C, so a 1 °C error is a 2 % error.
  • CO₂ absorption. Ultrapure water exposed to air jumps from 0.055 to over 1 µS/cm within minutes.
  • Polarisation and fouling. Electrodes coated with oil or scale read low.
  • Probe spacing and sample size. Small wafers need correction factors.
  • Moisture on insulators. Humidity can change surface resistivity by decades.

What goes in the uncertainty budget

  • Standard solution uncertainty and its temperature dependence.
  • Temperature measurement and compensation model.
  • Cell constant stability and fouling.
  • Geometry for wires and bars (length, diameter) and probe spacing for wafers.
  • Resolution and repeatability of the meter.

Combine the contributions as in the uncertainty budget calculator, then decide pass or fail with a guard band as explained on the calculators page.